The numbers a procurement officer should hold in their head
Three figures, drawn from the WHO prequalification standards and the peer-reviewed cold-chain literature, anchor most field decisions in this hub. They are worth committing to memory before any catalogue is opened.
Three cloudy days of autonomy
Every WHO PQS-prequalified solar direct-drive refrigerator is required to demonstrate a minimum autonomy of three consecutive cloudy days — the number of sunless days it can hold +2 to +8°C with no solar input. High-performing units must achieve this while rated to operate at ambient temperatures up to +43°C, and PQS defines holdover as the hours all points in the vaccine compartment stay below +10°C at that rated maximum ambient. A brochure 'holdover' quoted at a temperate 25°C is not the number you are buying; the PQS figure at your climate zone is.
±0.5°C, logged every minute
A PQS E006 30-day electronic refrigerator logger reads to an accuracy of about ±0.5°C with a minimum logging interval of one minute. That resolution is what separates a real excursion from sensor noise, and it is why a hand-read dial — accurate to a degree or two, sampled twice a day — cannot substitute for it. The logger does not make the refrigerator colder; it makes the failure visible.
14–35%, and 75–100%
The most-cited figures in this hub come from a systematic review of the published cold-chain evidence (Matthias et al., 2007): individual studies found freezing exposure in 14% to 35% of the refrigerators or shipments they examined, but studies that followed the whole distribution chain found freezing somewhere along it in 75% to 100% of chains. More rigorous study designs found more freezing — the strong signal that freeze exposure is under-detected, not over-stated. This is the single statistic that should reframe a procurement officer's instinct from 'keep it cold' to 'keep it in range, and watch the cold end hardest'.