What the E006 specification actually guarantees
A logger's value rests on a small set of performance properties the WHO PQS E006 standard defines and tests. Knowing what the standard does and does not promise is what separates an informed procurement from a brand-led one.
Accuracy and resolution
A PQS E006 30-day refrigerator logger reads cabinet temperature to an accuracy on the order of ±0.5°C with a minimum logging interval of one minute. That resolution is the point: it is fine enough to distinguish a genuine excursion from sensor noise, and frequent enough that an overnight dip is captured rather than averaged away. A twice-daily manual dial read, accurate to a degree or two and sampled twelve hours apart, simply cannot see the event the logger is built to catch.
On-device history, not just a download
The standard requires the device to display its recent status and alarm history on its own face. This is the human-factors heart of the specification — a monitor that can only reveal an excursion through a cable, a laptop, and trained software is, in field practice, the monitor that goes unread. The 30 days of on-screen history is what makes a daily check a glance rather than a procedure.
Validated alarm logic, pre-set
E006 devices ship pre-configured with heat and freeze alarm thresholds that combine temperature with duration. The freeze alarm triggers on a brief excursion below 0°C because freeze damage to adjuvanted vaccines can be near-instantaneous and irreversible; the heat alarm tolerates a short spike because heat damage is cumulative. The standard validates this logic so a field team does not have to configure it — and so loosening it in the field is recognisable as the error it is.
What the standard does not cover
E006 certifies the device, not the deployment. It cannot guarantee the logger is read, that an alarm is answered, or that the battery has been replaced before it died. Those are programme responsibilities the specification is silent on — and, as the field record repeatedly shows, they are where monitoring actually fails.