The WHO sequential testing strategy
WHO's consolidated testing guidance is built around the sequential use of up to three different assays to reach a final HIV-positive diagnosis, with national algorithms specifying which products fill each position. The logic is asymmetric, and that asymmetry is the whole point.
The structure
- A single non-reactive first test is a final negative in the testing context — subject to the window-period caveat below. One clear non-reactive result ends the algorithm.
- A reactive first test triggers a second, different assay. The tests are chosen from different manufacturers or different antigen designs so that a shared failure mode does not carry through.
- Reaching a positive diagnosis requires the full sequence of reactive results — under current WHO guidance, three consecutive reactive tests for a final positive — to drive the overall positive predictive value to the recommended 99% or higher.
- Discordant results are resolved by the algorithm, not by clinician preference: a defined repeat or referral pathway handles a reactive-then-non-reactive sequence rather than leaving it to judgement.
Product selection thresholds
WHO guidance frames minimum performance for the assays that fill the algorithm: a first-line test with high clinical sensitivity (on the order of ≥99%) so true positives are not missed, and downstream tests selected for high specificity (≥98%, with ≥99% expected of later-line assays) so false positives are filtered out. The first-line position is chosen to catch infection; the confirmatory positions are chosen to exclude error. Procurement to WHO prequalification status, where available, is the baseline — and the algorithm is validated as a whole, not assay by assay.